3-14
Displaying the NIBP Table
The Prism and Prism SE monitors stores up to 100 NIBP readings, as well as any other
parameters being monitored, with their corresponding time of measurement. When you turn the
NIBP table ON, the most recent readings will be displayed in tabular format in the NIBP trace
(NIBP must be set up in one of the three lower traces in the waveform display zone).
Additionally, tabular information for all other parameters being monitored will also display in
their corresponding times. You can press the RECORD key followed by the NIBP softkey to
print all readings stored in memory.
Invasive Blood Pressure
INVASIVE PRESSURE monitoring allows user to simultaneously derive pressures from up to 3
different transduced lines. The system works with standard pressure transducers having
5 µV/V/mmHg sensitivity. Following placement of invasive lines, evaluate alarms and alarm
limits to insure settings appropriate for patient. Default settings are set to label most commonly
used pressures.
1. For optimum waveform size on screen, connect higher pressures to BP1 and lower pressures
such as pulmonary artery, CVP, or ICP to BP2 and BP3.
2. Connect transducer cable to monitor. Invasive BP automatically activates on screen.
To display IBP in Trace slot:
1. Press the SYSTEM SETUP key to open the System Setup menu.
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
2. Press the TRACE SEL softkey to open the following menu.
DISP
VSC
DISP
OCRG
TRACE
2
AUTO
TRACE
3
AUTO
TRACE
4
AUTO
MORE
3. Scroll through TRACE 2, 3, or 4 to highlight desired IBP.
4. Press the PAGE HOME key to initiate change.
5. Insure pressure line is level with transduced vessel or its reference point.
6. Open the BP setup menu by pressing the softkey next to the BP label:
ALM
ON
OFF
ALM
LIM
SIZE/
SCALE
SITE
OFF
ART
ZERO
ALL
ZERO
BP1
MORE
7. Press the ALM ON softkey to activate alarms.
8. Press the ALM LIM softkey to set alarms for displayed values.
9. Press the SIZE/SCALE softkey to change displayed scale or to turn scale on/off. A new
scale for the pressure line can be set at any time. Changing scales may be necessary to insure
an appropriate waveform display.
Komentáře k této Příručce