
9-22
Table 5: Remote Status
Remote Status
Hex Value Description
Null Status 00h No activity
Low Battery 10h Battery voltage is low
Alarm Suspend 70h Alarm Suspend activated
NOTE: In the event a low battery condition occurs coincident with
an alarm suspend, the monitor will prioritize the event. Priority will
be given to the low battery signal and the CIS module will transmit
10h.
Table 6. Ports 1, 2, & 3 Backport Waveform Bit Values
MSB LSB
7 6 5 4 3 2 1 0
Port 1 Port 2 Port 3
Ports 1 Waveform Bit Values
B7 B6 Waveform
0 0 Invalid Selection
0 1 ECG Waveform
1 0 Invalid Selection
1 1 Invalid Selection
Ports 2 and 3 Waveform Bit Values
B5/2 B4/1 B3/0 Waveform
0 0 0 Invalid Selection
0 0 1 BP1 Scale Mode
0 1 0 BP1 Pulse Mode
0 1 1 BP2 Scale Mode
1 0 0 BP2 Pulse Mode
1 0 1 Respiration
1 1 0 SpO
2
1 1 1 CO
2
Komentáře k této Příručce