Escort Prism 20401 Specifikace Strana 113

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 170
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 112
8-10
3. Number “2”—select the energy to be delivered using the ENERGY SELECT knob on the
front panel.
NOTE: If the energy select knob is changed after a charge is
initiated, the energy will change to the new level immediately, and an
intermittent tone will sound during the change. Once it has reached
the new level, the tone will sound continuously.
4. Number “3”—press the CHARGE button on the defib/pacer front panel. The charge
indicator will flash with an intermittent tone until the energy reaches the pre-selected level,
when you will hear a continuous tone and the light stays on.
NOTE: If the indicator does not illuminate and/or a warning beep
sounds, check the paddle cable connection on the back of the
defib/pacer (behind the monitor).
NOTE: An intermittent warning tone will sound about 30 seconds
after being charged, and the stored energy will automatically
discharge internally after approximately 60 seconds if the unit is not
manually discharged.
5. Instruct all personnel to stand away from the patient and the bed before proceeding.
6. Discharge the defibrillator by simultaneously pressing and holding all four buttons on the
“hands-off” adapter cable’s junction box.
Zobrazit stránku 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 169 170

Komentáře k této Příručce

Žádné komentáře