8-2
connection, the unit must be operated by battery power.
CAUTION: Make sure to use an ECG cable with an inline resistor so
the monitor is protected from high voltage during defibrillation.
CAUTION: Do not operate defibrillator in the presence of high-
energy electrical fields. External electrical noise may corrupt the
ECG tracing.
CAUTION: Do not immerse paddles, cables, or electrical connector
in liquids during cleaning and disinfecting. Do not allow liquid or
electrode gel to enter switches, cable entrance, or connector.
Symbols (Defib/Pacer)
The following cautionary symbols appear on the defibrillator/pacer. Familiarize yourself with
these symbols and their meaning before using the defibrillator/pacer.
Type CF Defibrillation Protected Equipment: Isolated patient connections
comply with allowable leakage current limits for direct cardiac application and
are protected against effects of defibrillation.
Attention: Consult accompanying documents.
Replace Fuses as Marked: For protection against fire, replace only with fuses
of same type and rating.
Caution: Dangerous voltage.
Alternating Current: 100-240V @ 50/60 Hz.
Fuse: Voltage and current ratings appear adjacent to symbol.
Cleaning, Maintenance, & Calibration
Clean the Defibrillator/Pacer as required per hospital procedures.
There are no user-serviceable parts in the Defibrillator/Pacer. All maintenance and calibration
should be performed by qualified personnel only. Circuit diagrams, component part lists,
descriptions, and calibration instructions are available on request. Storage and transport
specifications are: Temperature = -10 degrees Celsius to 40 degrees Celsius, Humidity = 0% to
90%, Pressure = 500 to 1060hpa.